Mindray iMEC-8Vet Portable Veterinary Monitor with EtCO2

$ 7245.00 Sale

Features

  • 8.4" High-resolution LED display with 8 waveforms
  • Portable, less than 3.2 kg
  • No-fan design for a quiet care environment
  • Easy to use for intuitive operation
  • Up to 4 hours continuous monitoring with large capacity Li-ion battery
  • Trust for veterinary use, parameters including ECG, RESP, SpO2 , NIBP, PR, TEMP, IBP, CO2
  • 360-degree visible alarm indicator
  • 120 hours trend and 48 hours waveforms reviewing

Standard Configuration

  • ECG
  • RESP
  • NIBP
  • Mindray SpO2
  • TEMP
  • Li-ion battery

Specifications

  • Weight: ≤3.2 kg (8.5 lb)
  • Size: <210 x 270 x 112 mm (8.3 x 10.6 x 4.4 inch)
  • Display: 8.4" high-resolution LED with 800 x 600 pixels
  • Waveforms: Up to 8 traces
  • ECG: 5 or 3 lead: I, II, III, aVR, aVL, aVF, V
  • Respiration: RA-LA/RA-LL Impedance
  • SpO2: Masimo SET, Nellcor Oximax
  • NIBP: Manual, Auto, STAT
  • Temperature: 2 channels
  • EtCO2: Sidestream CO2 or Mainstream CO2
  • Recorder: 3-trace thermal array, 50 mm
  • Reviewing: 
    • 120 hours trends
    • 100 alarm events and associated waveforms
    • 1000 groups NIBP
    • Up to 48 hours waveforms
  • Download to PC: USB
  • Video Output: 1VGA
  • Wired Network: 1 RJ45

ECG

  • RR Range: 15 - 350 bpm
  • Speed: 6.25/12.5/25/50 mm/s
  • ECG Gain: 1.25/2.5/5/10/20/40 mm/mV, Auto Sweep
  • Mode: Diagnostic, Monitor, or Surgical

PR Range

  • 25 to 240 bpm (Masimo SpO2)
  • 20 to 300 bpm (Nellcor SpO2)
  • 25 to 350 bpm (IBP)
  • 40 to 240 bpm (NIBP)

NIBP Range

  • Systolic 40 to 270 mmHg
  • Diastolic 10 to 210 mmHg
  • Mean 20 to 230 mmHg
  • Interval in Auto Mode:
  • 1/2/2.5/3/5/10/15/20/30/60/90/120/180/240/ 480min
  • Max Time Allowed: 180 s

Respiration

  • Lead: I, II or Auto
  • RR Range: 0 to 150 rpm
  • Accuracy: 7 to 150 rpm: ±2 rpm or ±2%, whichever is greater

Nellcor Oximax SpO2

  • Measurement Range: 20-300 bpm
  • Accuracy:
    • 20 to 250 bpm: ±3 bpm
    • 251 to 30 bpm: not specified

Masimo SET SpO2

  • Measurement Range: 25-240 bpm
  • Accuracy:
    • ±3 bpm (measured without motion)
    • ±5 bpm (measured without motion)

C.O.

  • Range:
    • C.O.: 0.1 to 20 L/min
    • TB: 23 to 43°C
    • TI: 0 to 27°C
  • Accuracy:
    • C.O.: ±5% or ±0.1 L /min whichever is greater
    • TB, TI: ±0.1°C (without sensor)
  • Resolution:
    • C.O.: 0.1 L/min
    • TB, TI: 0.1°C

IBP

  • Range: -50 to 300 mmHg
  • Accuracy: ±2% or±1mmHg whichever is greater

EtCO2 - Sidestream

  • Sample Rate: 70 /100 ml/min
  • Range: 0 to 99 mmHg
  • awRR Range: 0 to 120 rpm
  • Accuracy:
    • 0 to 40 mmHg, ±2 mmHg
    • 41 to 76 mmHg, ±5% of the reading
    • 77 to 99 mmHg, ±10% of the reading

EtCO2 - Mainstream

  • Range: 0 to 150 mmHg
  • awRR Range: 0 to 150 rpm
  • Accuracy:
    • 0 to 40 mmHg, ±2 mmHg
    • 41 to 70 mmHg, ±5% of the reading
    • 101 to 150 mmHg, ±10% of the reading
    • 71 to 100 mmHg, ±8% of the reading
Go to full site